Geomet home
Stylus Manager Sub Menu
Search tool Information index
 
       
7.01.029 Geomet Junior+, 101, 101+, 301, 501 
      pre-version 7.01
  Definition
 
By default, Geomet sets the number of hits required for a Qualification to 5. This may not be suitable for your application if you are working with long styli or critical dimensions.

Stylus Manager Qualify
  Step 1
 
To change the default number of hits for the current session of Geomet, right-click on the Qualify dialog and a small sub-menu appears.

Choose <Set Default Number of Points>.

Set Number of HitsAlternatively you can left-click on the graphic of the stylus.

Stylus Manager Qualify
  Step 2
 
Enter the number of Points for your Qualification process. This new value is written into the Geomet System Options database and becomes the new default value. Enter number of Points
  Geomet System Options
 
The default number of Points required during the Qualificaton process is maintained in the Geomet System Options database. To change the default value, launch the System Options:

Keyboard: <F9>
Menu: [System→System Options]

Filter the System Options display be typing "Reference" in the Quck Search cell. Click on the listing:

"Number of data points used during calibration..."

Enter a new value in the pop-up tool. Press <Ok> twice to save the chnages.
System Options
  Stylus Manager, Features, Menus and Support Tools