USB Joystick

Back

 
 
 

Site Map

 

helmel.com

Introduction

Series 1 and 2 Helmel Joysticks, require calibration through the Windows Control Panel. If you are not sure of which Series Joystick you have, check here.

Windows recognizes the Helmel CMM Joystick as a standard 5 button USB gaming device. When you plug the Helmel Joystick in for first time, Windows will acknowledge it has identified new hardware and if required will install the driver.

Verifying Joystick Installation

To launch the Windows Device Manager, locate your Windows Control Panel (on XP, Start Menu→Control Panel) and locate "System". The System Properties tabbed dialog appears which you would choose the "Hardware" tab. Next select the <Device Manager> button and the Device Manager tool appears, see figure 1.

JoystickDM.jpg (30736 bytes)

figure 1, Windows Device Manager

The installation of the Helmel Joystick will be found under "Human Interface Devices" as shown in figure 1. Should the Device Manager indicate there is a problem with the installation, a "?" will appear next to the suspect device. If this happens try moving the USB cable to another port on you computer.

Calibrate the Joysticks

To calibrate the joystick movements, use the Windows Control Panel to locate "Game Controllers, see figure 2.

CalibJoy1.jpg (20233 bytes) CalibJoy2.jpg (20507 bytes)
figure 2, Control Panel figure 3, Game Controllers

Under Game Controllers an entry for the Helmel CMM Joystick is noted, see figure 3. Select <Properties>, see figure 4.

CalibJoy3.jpg (20269 bytes) CalibJoy4.jpg (18043 bytes)
figure 4, joystick Properties figure 5, Game Controller Calibration

To calibrate start by selecting <Settings> from the property page, see figure 5. Select <Calibrate>, a welcome page is displayed, select <Next>. The following page prompts you to leave the joysticks in the center location and press any button on the joystick, see figure 6.

CalibJoy5.jpg (11022 bytes) CalibJoy6.jpg (13061 bytes)
figure 6, Start Calibration figure 7, perform Calibration

When the Calibration Wizard acknowledges the button press, the next page will be displayed. This page prompts the operator to rotate the XY axis joystick in complete circles at the full extent, see figure 7. After a moving the XY joystick in complete circles, release the joystick and press any button. The Calibration Wizard will proceed to calibrate the Z axis in a similar process followed by calibrating the R axis. Follow all prompts until the Calibration Wizard has completed.

You will be returned to the Joystick Properties page where you should press the <Apply> button to save the calibration results.